Extending the Lifetime of Your Reverse Phase Columns: Cleaning and Regeneration Techniques

  • MS
  • Apr 3
  • 6 min read

Reverse-phase columns are essential tools in analytical chemistry, but various factors can cause their performance to deteriorate over time. Proper cleaning and regeneration techniques are essential for maintaining column efficiency and prolonging its lifespan.

 

This blog is part of our two-part series on extending the Lifetime of reverse phase high performance liquid chromatography (HPLC) columns. Our previous blog explores the probable causes of column contamination. This blog provides an overview of column cleaning and regeneration techniques.

The figure is an illustration of steps to clean, extend, and regenerate an  HPLC column to extend its lifetime.
Figure 1. Illustration of steps to clean, extend, and regenerate an HPLC column to extend its lifespan.

 

Below, we discuss steps such as sample preparation, mobile phase considerations, column care, storage, and regeneration.


1) Sample Preparation. 

Proper sample preparation is crucial for successful HPLC analysis. It directly impacts the accuracy, precision, and longevity of your column. Ensure that samples are adequately filtered and free of any particulate matter.


  • Filtration. Always filter samples through a 0.45 µm or smaller filter. This aids in removing particulate matter (such as dust, debris, or insoluble matter) that can clog the column bed and damage the delicate stationary phase, leading to increased backpressure and a reduced flow rate.


  • Dilution. Dilute the sample to bring its concentration within the detector’s linear range. This dilution not only helps to improve peak shape and resolution, but also reduces the risk of overloading the column.


  • Sample Clean-up. If your samples contain high levels of interfering compounds (such as proteins, lipids, and salts), consider using clean-up techniques such as solid-phase extraction (SPE), liquid-liquid extraction (LLE), protein precipitation, or any appropriate technique.


By diligently performing proper sample preparation, you can protect your HPLC column by minimizing the risk of column contamination


2) Mobile Phase Considerations.

  • Purity. High-purity solvents (HPLC or LC-MS grade) should be used to minimize the introduction of impurities. Utilize a high-quality water purification system, such as reverse osmosis or distillation, to effectively remove organic matter, ions, and microorganisms.


  • Degassing. Degassing removes dissolved gases (such as oxygen and nitrogen) to prevent bubble formation, which is essential for maintaining a stable and reproducible flow of mobile phase through the column. A stable mobile phase flow is fundamental to achieving accurate, reliable, and reproducible chromatographic results.


  • Buffer Preparation. Prepare buffers fresh to minimize microbial growth. Filter buffers to remove particulate matter.


  • Clean solvent bottles. Ensure that the glass bottle used for mobile phase storage is clean and free from contaminants or water remnants. This cleaning minimizes the risk of column contamination, prevents the growth of microorganisms, and ensures the reliability of your HPLC analysis.


3) Column Care.

  • Use Guard Columns. Place a guard column filled with the same packing material as the analytical column before it. This protects the analytical column from sample contaminants. The guard column, packed with the same material, retains compounds before they reach the analytical column. This effectively extends the analytical column's lifespan by preventing these contaminants from fouling or damaging the main separation bed. Replace the guard column regularly to maintain its effectiveness.


  • Employ Proper Flushing Procedures. Flushing an HPLC column is critical in maintaining its performance and extending its lifespan. The manufacturer's recommendations for flushing procedures should always be followed as they are specifically tailored to the column's unique characteristics. These procedures ensure that the column is effectively cleaned of residual contaminants without damaging the stationary phase. Different stationary phases may require different flushing solvents and procedures. The length and internal diameter of the column can influence the flushing time and solvent volume. Proper flushing helps to maintain consistent column performance, including peak shape, retention times, and efficiency.


    Incorrect flushing procedures can lead to:

    • Chemical degradation. Harsh solvents or inappropriate pH conditions may degrade the stationary phase. Use appropriate solvents to remove residual contaminants.


    • Physical damage. Aggressive flushing conditions can damage the column packing.


Typical Flushing Steps.

  • Remove Residual Mobile Phase. Flush the column with several column volumes of a suitable solvent to remove the previous mobile phase. This often involves a gradual increase in the concentration of organic solvent.


  • Remove Buffers and Salts. If buffers are used in the mobile phase, flush the column with water or a dilute acid or base solution to remove salt deposits.


  • Remove Strongly Retained Compounds. Stronger solvents or specialized cleaning solutions may be necessary for compounds that are strongly retained.


  • Equilibrate with Starting Mobile Phase. After flushing, equilibrate the column with the starting mobile phase for several column volumes to ensure stable baseline conditions.


  • Store Columns Properly. Storing HPLC columns correctly is crucial for maintaining their performance and extending their lifespan. Whenever possible, store the column in its original packing.


    Points to consider while storing columns include

    • Storage Solvent. Columns should be stored in the storage solvent recommended by the column manufacturer. This solvent is specifically chosen to maintain the integrity of the stationary phase. Acetonitrile or Methanol are widely used for storing reversed-phase columns. Some columns may require different storage solvents, such as isopropanol or specific mixtures.


    • Temperature. Store columns at room temperature or as recommended by the manufacturer. Freezing can damage the column packing.


    • Environment. Store columns in a clean, dry, and dust-free environment. Humidity, along with any residual water in the column from the stored mobile phase, can also significantly impact column integrity during storage. Store columns securely to prevent physical damage.

 

  • Column Preparation for Storage.

    • Flush Thoroughly. Before storage, flush the column with several column volumes of the recommended storage solvent. This removes any residual mobile phase components or contaminants.


    • Ensure Complete Flushing. Ensure that all traces of the previous mobile phase are removed. Incomplete flushing can lead to precipitation of salt or microbial growth within the column.


    • Column Capping. Always cap the column ends when not in use to prevent the stationary phase from drying out, which can cause irreversible damage to the column packing. Capping the column will also protect the column against dust and other airborne particles.


  • Regeneration of HPLC Columns. Regeneration is a more intensive cleaning process than routine flushing. It's typically performed when column performance has significantly deteriorated due to severe contamination.  


    When to Consider Regeneration?

    • Significant increase in backpressure. This indicates a substantial blockage within the column.


    • Severe peak tailing or broadening. These issues suggest strong interactions between analytes and contaminants within the column.


    • Loss of resolution. A significant decrease in peak separation indicates a loss of column efficiency.


    • Ghost peaks. Persistent peaks that appear even in the absence of injected analytes suggest the presence of strongly retained contaminants.


Regeneration procedures vary depending on the nature of the contamination and the type of column. Some general steps include:  

  • Removal of Residual Mobile Phase. Flush the column with several column volumes of a suitable solvent to remove the previous mobile phase.  


  • Removal of Buffers and Salts. Flush the column with water or a dilute solution of acid or base to remove salt deposits.  


  • Removal of Strongly Retained Compounds. This may involve using stronger solvents, such as:

    • Acetonitrile/Water mixtures. For moderate levels of contamination.


    • Methanol/Water mixtures. For more stubborn contaminants.


    • Isopropanol. A strong solvent that can remove strongly retained compounds.  


    • Specialized cleaning solutions. Some manufacturers offer specific cleaning solutions for their columns.  


  • Removal of Particulate Matter. If suspected, backflush the column to remove particulate matter trapped at the inlet frit.  


  • Equilibration. After the cleaning process, equilibrate the column with the starting mobile phase to ensure stable baseline conditions.  


Always consult the manufacturer's instructions for specific regeneration procedures and recommended solvents. Use caution when using strong solvents, as they can damage the stationary phase if used improperly. Monitor the backpressure and flow rate during the regeneration process. After regeneration, evaluate column performance by injecting a standard sample and assessing peak shape, retention times, and resolution.  


Regeneration may not always be successful in restoring a severely contaminated column to its original performance. In some cases, column replacement may be necessary. By following these guidelines and consulting the manufacturer's recommendations, you can effectively regenerate your HPLC columns and extend their useful life.


In summary, by understanding the causes of contaminant buildup and implementing appropriate preventive measures, you can significantly extend the lifetime of your reversed-phase HPLC columns and maintain reliable chromatographic performance. By adhering to these proper storage practices, you can significantly extend the lifespan of your HPLC columns, ensure consistent performance, and ultimately reduce the overall cost of analysis.


For more information on how our NanoPak-C All Carbon media or custom media services can address your pharmaceutical purification challenges, or to request samples, please email us at inquiry@millennialscientific.com, call us at 855-388-2800, or complete our online form.

Comments


Commenting on this post isn't available anymore. Contact the site owner for more info.

© Millennial Scientific, Inc. 2019

 sales@MillennialScientific.com

www.MillennialScientific.com

MBE%2Blogo%20(1)_edited.png

Seller Credentials:

Registered Small Business

MBE - Certified Minority Business Enterprise

DBE - Certified Disadvantaged Business Enterprise

ISO 9001 2015 Quality Management Systems

ISO 27001 2017 Information Security Management

StartUp NY Company

Long Island Innovation Hot Spot Company

iso-9001-2015-standard.jpeg
bottom of page